What You See on the Display What You Do Comments
7-4
Questions? Refer to Page 10-10 for Technical Support Instructions.
12. Follow the instructions in the CVC kit
package insert. Then apply a drop of the
sample to the target area.
When sufficient sample has been applied,
the monitor beeps, displays Sample
Accepted and automatically starts the test.
For a detailed description of the test strip
target area, refer to the Test Strips section of
Chapter 1.
13. Wait for the monitor to analyze the
sample and display the test result.
The monitor counts down then displays
the test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
14. Note the result. If required, scan or
manually enter the comment code, and
press Enter.
If there is no prompt to enter a comment
code, skip down to step 15.
The monitor may be enabled to scan or
enter a 1- to 2-digit comment code.
15. You can select one of the following
options:
Press 1 – New Level.
Press 2 – Same Level.
Press the Menu button to return to the
Menu Mode.
Press On/Off to turn off monitor.
Remove the test strip from the monitor
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
The display shows:
Level (test level 1 through 5)
Test (replicate 1 through 4 for that level)
Comment Code (Number in box)
Date and Time of test
Operator ID (o: prompt)
Note that test date and time alternates with
Operator ID.
If 2 is selected, the Scan or Enter Strip
Lot screen will appear. Return to step 8 to
continue linearity testing with the Same
Level.
If 1 is selected, the Select Level screen will
appear. Return to step 7 to perform linearity
tests for a New Level.
45 mg/dL
45 mg/dL
Chapter 7: Linearity Test