Abbott Precision Xtra Bedienungsanleitung Seite 56

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 60
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 55
52
52
CHAPTER 9
Caring For Your Meter
Questions?
Call Customer Service:
1-800-527-3339
Cleaning Your Meter
Store your meter in its carrying case. Avoid getting dirt,
dust, blood, control solution, or liquid in the meter test
strip port. Clean the outside of the meter using a damp
cloth and mild soap.
Healthcare professionals: Acceptable cleaning
solutions include:
•70%isopropylalcohol,or
•Amixtureof1partammonia,9partswater,or
•Amixtureof1parthouseholdbleach,9partswater
IMPORTANT:
Do not try to clean the strip port.
Do not pour liquid into the strip port or onto the buttons.
Do not immerse the meter in water or any other liquid.
Replacing Your Meter’s Battery
When your meter needs its batteries replaced,
the display window shows one of two things:
This means the battery is low. You
may still use your meter and the results
will be accurate. However, the backlight is not
useable. It is recommended that you replace
the battery at this time.
This means the battery must be replaced.
The meter is not useable. The meter turns off
automatically.
IMPORTANT:
Do not remove the old battery until you have
a new battery to install. It may be necessary to reset the
time and date once you install a new battery.
Caring for
Your Meter
Seitenansicht 55
1 2 ... 51 52 53 54 55 56 57 58 59 60

Kommentare zu diesen Handbüchern

Keine Kommentare