37
5
To mark the test as a control solution
test, Press and Release the
button
once.
shows on the display
window.
IMPORTANT: If you do not mark the test as a control
solution test, it will be saved in your meter’s memory as one
of your personal blood glucose or blood ketone results. This
could affect your blood glucose averages.
Applying Control Solution to the Test Strip
1
Turn the control solution bottle upside down three to
four times to mix the solution.
2
Remove the cap.
3
Apply a drop of control solution to
the white area at the end of the test
strip. The control solution is drawn into the test strip.
4
Continue to touch the control solution drop to the
test strip until the meter begins the test.
The meter begins the test when:
•You hear the beeper, if the beeper is ON.
•The display window shows the status bar.
•Then the display window shows the countdown.
Status bar
Countdown
Control Solution
Testing
Kommentare zu diesen Handbüchern