Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
18
The Operator ID just entered is not on the
list of acceptable Operator ID’s defined
for this monitor.
1. Testing cannot continue because
the Operator ID is not certified.
2. Contact nursing director for
certification guidance.
The Operator ID entered has expired.
(The menu item 2-Continue appear only if
this option is set to Warn.)
1. Contact nursing director for
recertification directions.
The operator ID entered expires in the
shown number of days.
1. Press 1 to continue with testing.
2. Run two levels of QC correctly
before the number of days
passes to maintain competency
and access to the meter.
The test result is below the measuring
limit of the system.
Press 2 – Repeat Test –
1. verify result by sending
sample to lab.
2. notify M.D.
The test result is above the measuring
limit of the system.
Press 2 – Repeat Test –
3. verify result by sending
sample to lab.
4. notify M.D.
The test strip or the test port is wet,
defective, contaminated, or the wrong test
strip was inserted.
Press 1 to continue testing with a
new test strip. Repeat test. If error
continues to appear, let meter sit for
24 hours and try again.
There may be a problem with the test
strip. One of several 4-digit error codes
may be displayed. These codes provide
Abbott personnel additional information
about the problem. Errors include:
4327 – The strip was removed during
testing.
4330 – Blood glucose may be too high to
be read by the system or there may be a
problem with the test strip.
1. Repeat test with a new test
strip.
2. If the error appears again,
confirm the result by sending a
sample to the laboratory.
3. Contact Abbott technical
support and give them the 4-
digit error code.
4. Contact POCT Coordinators.
Troubleshooting When a Control Test is Selected
Symptom/
Error Message
Explanation
How to Respond