Abbott Precision XceedPro Betreibershandbuch Seite 17

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 20
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 16
Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
17
The monitor is shipped from the factory
without configuration. This screen will
appear the first time you turn on the meter
after receiving it.
Contact the POCT Coordinators or
Nursing manager for replacement
meter.
QC Interval for Controls Exceeded.
Controls must be performed before
testing.
Press 1 to exit. Perform the control
tests indicated in the message.
The Upload Interval Exceeded. Press 1 to exit, or if it appears,
press 2 to continue testing. Place
the monitor in the docking station.
The room temperature is outside the
limits for testing.
Ensure that the room temperature is
within the specified limits. Press 1 to
exit.
The test strip was inserted at the wrong
time. The monitor will beep if functions
are attempted before the strip is removed
or if the test strip is left in when the test
calls for the test strip to be removed.
Remove the test strip from the test
strip port.
The scanned Patient 9-10 digit CSN is
similar in format to the test strip barcode.
1. Re-enter the patient 9-10 digit
CSN using the barcode scanner
or the keyboard.
2. Contact the POCT Coordinators
if the problem persists.
The barcode just scanned was not read
correctly. Possible reasons are:
The red scanner beam did not pass
across all the bars of the code
The barcode was damaged or poorly
printed
The barcode was not of the lot or type
specified for this ID or lot number.
1. Rescan the barcode.
2. If problem persists, contact the
POCT Coordinators.
The lot number just scanned or entered is
not on the list of acceptable lot numbers
defined for this monitor.
1. Verify and re-enter the lot
number using the barcode
scanner.
2. Manually enter the ID or lot
number.
3. If the problem persists, notify
the POCT Coordinators.